73
e. ACCEPT? YES/NO: If necessary, use the arrow keys to move to any field and change the
data entered. To abort this screen entirely, press NO at this prompt. Any data entered on this
display will be aborted and the display will return to the options menu . If the parameters
entered are correct, press the YES/ENTER key, and the new rate or rates will be accepted. The
display returns to the Options menu when this screen is exited.
f. New values will be in effect when the Options menu is accepted.
7 DN Occlu (Down Occlusion): This feature allows the pump to sense an occlusion pressure
in the line between the pump and the access site, e.g., a kink in the tubing between the pump
and the patient, a blockage in the access site, or a clamp left on the distal tubing. There are two
options for this field: the “Low” setting triggers the alarm when it detects in-line pressures at 400
mmHg ±150 (8 ±3 psi “pounds per square inch”), or the “High” setting (default setting) alarms
when it detects in-line pressures at 900 mmHg ±150 (18 ±3 psi). Press the NO/CHANGE key
to change the selection. When the correct pressure setting appears, press the YES/ENTER key to
accept it and move the cursor to the next field. If setting is changed, the new settings will remain
in effect until a new therapy is programmed.
8 AIR SENS (Air In Line Sensor): This feature allows you to select the amount of air the pump
will detect in the administration set tubing before it goes into an alarm state. The choices are 0.1
mL, 0.5 mL (the default setting), 2 mL, and OFF. Press the NO/CHANGE key to change the
options until the desired setting appears on the display, and then press YES/ENTER to select
and move the cursor to the next line. If setting is changed, the settings will remain in effect until
a new therapy is programmed.
Whenever the Air-In-Line Sensor is set to OFF, Moog Curlin Infusion
recommends the use of an air-eliminating filter.
When the Air-In-Line Sensor is set to “OFF” and the RUN key is pressed at the Run Option
screen, the alert message below will appear on the screen. It will also appear if the Air-In-Line
is changed to “OFF” from the OPTIONS menu when an infusion is in progress and the AIR
SENS field is accepted.
A i r - I n - L i n e S e n s o r O f f M e s s a g e S c r e e n
This message alerts you to the Air-In-Line Sensor being set to “OFF.” The message will require
answering “YES” or “NO” to whether the administration set being used has an air-eliminating
filter. If you press the YES key to indicate that there is an air-eliminating filter in use, the Air-In-
Line sensor will be completely disabled and the pump will start the infusion.
IMPORTANT
INFORMATION
PCA T
her
ap
y
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...