95
Cause:
The pump has detected excessive positive pressure between the pump and the
IV Source.
Resolution:
Press PAUSE to clear alarm screen. Check for excessive pressure on the IV bag
and relieve it. When ready to begin infusion, select “RESUME.”
LCD Display:
***ALARM*** REPLACE SET 3
Alarm:
Replace Set
Cause:
The pump has detected that the disposable administration set must be
replaced.
Resolution:
Replace the disposable administration set, prime the set if necessary, and when
ready to begin infusing, select “RESUME” to continue the infusion.
LCD Display:
***ALARM*** REPLACE SET 4
Alarm:
Replace Set
Cause:
The pump has detected excessive pressure in the disposable administration set.
Resolution:
Make sure that all the IV clamps are open. Relieve pressure, open and close
pump door.
LCD Display:
***ALARM*** SET NOT INSTALLED
Alarm:
Administration Set Not Installed.
Cause:
The administration set has not been installed or has been installed improperly.
Resolution:
Install appropriate
Moog Curlin Infusion
Administration set for desired
therapy. Use ONLY
Moog Curlin Infusion
Administration sets. If the
blue tubing guide on the administration set has been damaged in any way,
replace the entire administration set. See Chapter 2, Figures 2.9 and 2.10 for
proper installation of administration set.
LCD Display:
***ALARM*** UNATTENDED PUMP
Alarm:
No key presses have occurred while the pump was paused
Cause:
When a therapy is in progress and a pump is placed in a paused state and
left unattended for more than two minutes, this alarm will sound. If the
prime menus are left unattended for one minute, this alarm will sound
Resolution:
Press the PAUSE key to clear the alarm screen. If required, return to the
screen displayed at the time of the alarm and complete the activity that was in
progress, or resume the infusion, or turn the pump off.
LCD Display:
***ALARM*** UP OCCLUSION
Alarm:
Negative pressure detected between the pump and IV reservoir.
Cause:
Pump has detected excessive negative pressure in the administration set
between the medication reservoir or IV bag (“up stream”) and the pump.
T
roubleshooting
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...