Resolution:
Recheck the field and accept it by pressing the YES/ENTER key.
LCD Display:
***ALERT *** GREATER THAN MAX FIELD LIMIT, MAX = _____
Alert:
Unable to program a field with a value that exceeds the maximum limit
Cause:
Value entered is greater than the allowable limit.
Resolution:
Recheck data entry. A value entered cannot exceed the field limit.
LCD Display:
***ALERT *** GREATER THAN WHAT BOLS INT ALLOWS
Alert:
Unable to enter a # of boluses/hr that exceeds the possible number allowed
given the bolus interval.
Cause:
# Boluses/hr value exceeds the possible number given the bolus interval
(“BOLS INT”) programmed.
Resolution:
Recheck data entry. The # of boluses/hr cannot be set for higher than the
number allowed using the given bolus interval.
LCD Display:
***ALERT *** INFUSION NOT IN PROGRESS
Alert:
Infusion not in progress.
Cause:
Modification attempted to Next Dose or Early Down Ramp field without an
infusion in progress.
Resolution:
Wait until infusion is in progress before modifying or modify original Rx
programmed.
LCD Display:
***ALERT *** INSUFFICIENT VOLUME MAX DOSES=_____
Alert:
Bag Volume value is insufficient to deliver the remaining
Cause:
Occurs whenever the prescription is changed so that bag volume is now less
than amount required for the entire therapy (based on volume already
infused plus volume required to deliver the remaining doses).
Resolution:
Recheck data entry. If the source container contains sufficient volume to
complete the remaining dose cycles (dose KVO volume), increase
the “Bag Vol.” If the volume is insufficient, decrease the “DOSE/BAG” such
that the therapy can continue with the new modified prescription.
Note:
If either “Bag Vol” or “DOSE/BAG” is modified and the therapy is
repeated, these fields will have to be verified against the source
container volume and prescribed prescription and changed
accordingly.
LCD Display:
***ALERT *** INVALID ENTRY
Alert:
Invalid Entry
Cause:
Data entered does not fall within the defined limits of the program.
Resolution:
Enter data that falls within the limits of the program. Refer to the specific
therapy section in the User’s Manual.
LCD Display:
***ALERT *** KVO RATE IS GREATER THAN INF RATE
T
roubleshooting
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...