98
ERROR CODES
If a pump goes into an error code, a numeric code will appear on the display screen. If this has
occurred, turn the pump off and back on. If the error code message does not reappear, continue
using the pump.
If the pump passes its start-up diagnostic tests, resume the therapy. If an error code recurs, turn
the pump off, replace it, and report the error code to the healthcare provider.
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
Turn pump Off/On
Air Sensor
Turn Pump Off/ON or Call Provider
Pressure Sens DN
Check Admin Set Turn pump Off/On
Pressure Sens UP Check Admin Set
Turn pump Off/On
Turn Pump Off/On or Call
Provider
Turn pump Off/On or Call
Provider
Turn pump OFF Call Provider
Call Provider
Turn Pump OFF Call Provider
Empty Lithium Battery Call Provider
Turn Pump OFF Call Provider
Turn Pump OFF Call Provider
Turn Pump OFF Call Provider
Turn Pump OFF Call Provider
Stalled Pump Turn Pump OFF/ON
Call Provider
Turn Pump OFF Call Provider
Turn Pump OFF Call Provider
Low Alarm Batt Connect AC Adapt
Turn pump OFF/ON
Turn Pump OFF
Call Provider
Turn Pump OFF
Call Provider
Turn Pump OFF
Call Provider
Low Alarm Audio
Connect AC Adapt
Turn pump OFF/ON
Turn Pump OFF Call Provider
Keyboard Buffer Full
Air In Line test circuit failure
Bad Down Strain Beam
Bad Up Strain Beam
Bad Encoder
Bad Analog to Digital Conversion
Bad Clock
Bad Air In Line comparator
register reading from FPGA.
Bad Door Sensor
Empty Lithium Battery
Bad Tristate Pin
Bad Test Pin
Bad Reverence Voltage
Bad Motor Supply
Broken Motor Shaft
Bad Motor Inhibit
Bad Motor Supply Enable
Low Auxiliary Battery
Bad Auxiliary Battery
Audio Main Speaker Low
Audio Main Speaker High
Audio Auxiliary Speaker Low
Audio Auxiliary Speaker High
Report Error Code to Factory
Report Error Code to Factory
Replace set and Resume
Replace set and Resume
Report Error Code to Factory
Report Error Code to Factory
Report error code to factory
Report Error Code to Factory
Report Error Code to Factory
Report error code to factory
Report Error Code to Factory
Report Error Code to Factory
Report Error Code to Factory
Turn power OFF/ON and resume
Report Error Code to Factory
Report Error Code to Factory
Connect Pump to AC Battery
Eliminator and Turn Power
OFF/ON to Pump. May Require
Turning Pump OFF/ON a Few
Times
Report Error Code to Factory
Report Error Code to Factory
Check/replace D31. Replace PCB
Connect Pump to AC Battery
Eliminator and Turn power OFF/
ON to Pump. May Require
Turning Pump OFF/ON a Few
Times.
Report Error Code to Factory
Error
Code
LCD Displayed
Message
Probable Cause
Corrective Action
10
Turn pump OFF/ON or Replace
Battery or Call Provider
Motor Control Loss
• Replace Batteries
• Connect AC Adaptor
• Turn power off/on and resume
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...