55
13. CHANGE CONTRAST: The contrast on the display screen is automatically optimized and is
no longer changeable with this option.
Changing Date and Time
14. CHANGE DATE AND TIME: When this field is selected, the screen on the left appears:
Change Date and Time Screen
This screen is used to set the system Real Time Clock in the pump and will change the date and
time displayed on the Welcome Screen.
Note:
The time must be reset to accommodate daylight savings time changes and also at any
time the pump is moved into a different time zone.
a. Time Format: You have the option of selecting a twelve-hour or twenty-four-hour time
format. Press the NO key until the correct selection appears on the field and then press the
YES/ENTER key to accept the selection and move the cursor to the next field.
b. AM/PM: The second field allows you to select AM or PM when the twelve-hour time format
is selected in the previous field. If the twenty-four-hour format was selected, this field will
not appear. Press the NO key until the correct selection appears on the field, and then press
the YES/ENTER key to choose the selection and move the cursor to the next field.
c. Time: Use the numeric keys to enter the correct time and then press the YES/ENTER key
to accept and move the cursor to the next field.
d. Using the numeric keys, enter the correct mm/dd/yy, using two digits for each (e.g., if the
month is January, enter 01; if it is a single digit date, use the number preceded by a 0 and
then enter the last two digits of the year code, 00, 01, etc.). When the date is correct, press
the YES/ENTER key to accept and move to the next field.
e. A final prompt will be given to “ACCEPT? YES/NO.” If any data needs to be corrected,
use the arrow keys to locate the correct field, press NO to change the data, enter the
correction(s), and then press YES/ENTER again to accept the correction. Move the cursor
back to the “ACCEPT? YES” field and press YES/ENTER. This action enters the new time
and date and moves to the next field in the BioMed Setup Menu. Pressing the NO key on
this field aborts the change date and time selection and returns to the BioMed Setup Menu.
Basics Pr
og
ramming
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...