41
8.
Percent Bolus (%Bols):
This screen provides Bolus Volume as percentage of Total
Volume. TBOL is the total volume infused through bolus, TVOL is the total volume
infused, and % BOLS is the percentage of TBOL over TVOL. Changes to this screen are
not allowed.
9 .
Clear Shift Total (CLRShft):
Two Shift Total screens are displayed for 11 seconds
when this key is pressed. This key allows the user to view or CLEAR the Shift Totals (this
action does not interfere with the operation of the pump).
To
CLEAR
the Shift Totals information to zero, press the CLRShft/9 key, and then press
the YES/ENTER button in response to the “DONE? CLEAR” question and the pump
will go back to the running screen. Press the CLRShft/9 key again to verify that the values
are reset or cleared.
To
EXIT
the Shift Totals screens without clearing, press NO/CHANGE for the “DONE?
CLEAR” question and then YES/ENTER for “DONE? EXIT” and return to the running
screen. The pump will return to the running screen automatically after 11 seconds have
elapsed since the pressing of the CLRShft/9 key, even if no key is pressed.
10.
Check Maximum Infusion (ChckMax):
The information displayed here is the
maximum (rate or amount) that can be infused hypothetically, given the prescription
parameters. This “Maximum Infusion per Hour” serves as a “sanity check” after the pump
has begun infusing. Changes to this screen are not allowed.
8
%
BOLS
9
CLR
Shft
0
Chck
Max
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...