91
Resolution:
To terminate the current therapy in progress, press YES. Resume will
no longer be available. The Select Therapy menu will be displayed (if only
one therapy is in programmed, the pre Rx menu for the therapy will be
displayed). When the therapy to repeat is chosen, the last prescription data
programmed will appear in the menu, excluding Delayed Start and Load
Dose (if therapy was previously started). When the infusion is started the
entire therapy will be repeated.
To continue the current therapy in progress, press NO. The Pause menu will
be displayed.
LCD Display:
***ALERT *** Rx IN PROGRESS, YES New Program – NO to Return
Alert:
“NEW PROGRAM” was selected while a therapy is already in progress.
Cause:
“NEW PROGRAM” was selected and there is a therapy currently in progress.
Resolution:
To terminate the current therapy in progress, press YES. Resume will
no longer be available. The Select Therapy menu will be displayed (if only
one therapy is in programmed, an alert screen informing that the Rx will
be erased will be displayed). When the therapy for the new program is
chosen, entry of a new prescription will be required.
To continue the current therapy in progress, press NO. The Pause menu will
be displayed.
LCD Display:
***ALERT *** Rx Programmed, YES to Erase – NO to Return
Alert:
“NEW PROGRAM” was selected for a therapy that has an existing
prescription.
Cause:
“NEW PROGRAM” was selected for a therapy that has an existing
prescription.
Resolution:
To erase the prescription for the selected therapy, press YES. The pre Rx
menu will be displayed and entry of a new prescription will be required.
To preserve the prescription for the selected therapy, press NO. The Pause
menu will be displayed.
LCD Display:
***ALERT *** SET NOT INSTALLED
Alert:
Set Not Installed.
Cause:
Administration set is not properly installed.
Resolution:
Install the appropriate
Moog Curlin Infusion
administration set for the
desired therapy. Use
ONLY
Moog Curlin Infusion
administration sets. See
Chapter 2 figures 2.9 and 2.10 of the User’s Manual for proper installation of
the administration set.
LCD Display:
***ALERT *** SYSTEM CLOCK NOT SET
Alert:
System Clock Not Set.
Cause:
The clock for the pump is not set
T
roubleshooting
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...