71
S h i f t T o t a l s S c r e e n s
a. Cleared/ELAPSED: This field displays the date and time the shift total was last cleared. It
toggles with the amount of time elapsed since that clearing.
b. Vol INF/Amt INF (Volume/Amount Infused): This field displays the volume (mL) and
amount (mg or mcg) infused since the shift total was cleared.
c. BOLUS/Clin DOSE (Clinician Dose): This field displays the number of clinician doses
given since the total was cleared. It toggles with the number of patient activated boluses
received of those requested.
d. To exit the shift total feature without clearing the totals, press the YES key at the
highlighted EXIT field.
e. To clear the Shift Totals, press the NO key at the “EXIT” field. The field will change to read
“CLEAR,” press the “YES” key now to clear. A lock level 3 will require an access code entry
before allowing the user to clear the Shift Totals.
4 HRLY TOT (Hourly Totals): To view hourly infusion totals press the NO/CHANGE key
to change OFF to ON and then press YES/ENTER. Either of the following two screens will
appear:
H o u r l y T o t a l s S c r e e n
or
The screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a. The highlighted field has a marker indicating the current hour in progress.
b. “Volume” or “Amount” will show in the first column, “BG” (Bolus Given) in the second
column and “BA” (Bolus Attempted) in the third column.
c. Use the up arrow key to scroll back through the last 24 hours.
d. If the infusion was not running, a series of ------ will appear for that hour.
e. To exit this screen, press the NO/CHANGE key.
PCA T
her
ap
y
Summary of Contents for CURLIN INFUSION PainSmart
Page 24: ...21 Figure 2 8 Closing the Pump Door Pump Set Features...
Page 39: ......
Page 49: ......
Page 113: ......
Page 121: ......
Page 125: ...118...
Page 126: ...119 Technical Specifications...
Page 127: ...120...
Page 128: ...121 Technical Specifications...
Page 129: ...122...
Page 130: ...123 T...
Page 131: ...124...
Page 137: ......
Page 139: ......
Page 140: ......
Page 141: ...Rev I 360 9043...