DNA Engine Dyad Operations Manual
Opening a Program in Basic Mode
You can choose to open and edit a graphical program in either the graphical programming
window, or in the advanced programming window. Opening a graphical program in ad-
vanced mode is desirable if you wish to add step modification options, incubations below
0
°
C, or other programming features not available in graphical programs (see the “Types of
Programs” section in Chapter 5 for a listing of available program features). However, once
advanced-only features are added to a graphical program, the program can no longer be
opened or edited in basic mode.
We will open our example graphical program, GRAPH#1, in basic mode. The procedure is
similar to that described above for opening a program in advanced mode and is summa-
rized here.
• Select <Programs>.
• Select <Open>.
• Select <Basic Mode>.
The open program window will appear.
• Select the folder FOLDER1.
• Select the program GRAPH#1.
• Select <OK>.
GRAPH#1 will be displayed in the graphical programming window.
Editing a Program
Recall our general program.
Temperature Control Mode: Calculated
Lid Control Mode: Constant at 100
°
C
1. Incubate at 94
°
C for 00:01:00
Beep on Target (advanced only)
2. Incubate at 92
°
C for 00:00:30
Beep on Target (advanced only)
3. Gradient from 45
°
C to 65
°
C for 00:03:00
4. Cycle to step 2 for 29 more times.
5. Incubate at 10
°
C forever
End
Summary of Contents for DNA Engine Dyad PTC-220
Page 8: ...1 4 DNA Engine Dyad Operations Manual ...
Page 14: ...2 6 DNA Engine Dyad Operations Manual ...
Page 38: ...4 18 DNA Engine Dyad Operations Manual ...
Page 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Page 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Page 113: ...9 4 DNA Engine Dyad Operations Manual ...
Page 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Page 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Page 137: ...11 10 DNA Engine Dyad Operations Manual ...
Page 142: ......
Page 156: ...DNA Engine Dyad Operations Manual In 12 ...
Page 157: ...Declaration of Conformity DoC 1 ...