Specifications
Thermal range:
–5.0
°
to 105
°
C, but no more than 30
°
C below ambient
temperature (4
°
C to 105
°
C, but not more than 23
°
C
below ambient temperature for the Twin Towers
®
unit)
Accuracy:
+ 0.3
°
C of programmed target @ 90
°
C, NIST-traceable
Thermal uniformity:
+ 0.4
°
C well-to-well within 30 seconds of arrival at
90
°
C (for most Alpha units; see specifications for
individual Alpha units)
Ramping speed:
Up to 3
°
C/sec for all single- and dual-block Alpha units;
Up to 1.2
°
C/sec for the Twin Towers
®
unit
Sample capacity:
Varies with installed Alpha unit
Line voltage:
200-240VAC
Frequency:
50-60Hz
Power:
1600W maximum
Fuses:
Two 6.3A, 250V, 5 x 20mm
Displays:
One 1/4 size VGA screen (320x240), 16 colors
Ports:
One 9-pin RS-232 serial port
One ethernet port
Memory:
8 MB
Weight:
11kg ( base only)
Size:
48 x 29 x 15cm (l x w x h, base only)
Gradient Specifications (96 Alpha unit only)
Accuracy:
+ 0.4
°
C of programmed target at end columns,
30 seconds after the timer starts for the gradient step,
NIST–traceable
Column uniformity:
+ 0.4
°
C, well–to–well within column, within 30 seconds
of reaching target temperature
Calculator accuracy:
+ 0.4
°
C of actual well temperature
Lowest programmable
30
°
C
temperature:
Highest programmable
105
°
C
temperature:
Temperature differential
1–24
°
C
range for gradient:
Summary of Contents for DNA Engine Dyad PTC-220
Page 8: ...1 4 DNA Engine Dyad Operations Manual ...
Page 14: ...2 6 DNA Engine Dyad Operations Manual ...
Page 38: ...4 18 DNA Engine Dyad Operations Manual ...
Page 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Page 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Page 113: ...9 4 DNA Engine Dyad Operations Manual ...
Page 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Page 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Page 137: ...11 10 DNA Engine Dyad Operations Manual ...
Page 142: ......
Page 156: ...DNA Engine Dyad Operations Manual In 12 ...
Page 157: ...Declaration of Conformity DoC 1 ...