DNA Engine Dyad Operations Manual
10-6
Remote
Error Code
Error Code
Reason
Action
3
PSO-PT
The power supply temperature
was higher than normal. To
prevent instrument damage, the
program was automatically
stopped. Overheating may be
caused by inadequate airflow or
a problem with the power
supply.
Please observe the following
guidelines: 1) place instruments
at least 10cm apart, 2) avoid
placing instruments in any
location where hot air might
enter the intake vents, 3) place
instrument on a hard surface with
no debris or paper underneath,
4) clean all air vents and Alpha
unit fins of dust and debris, 5)
avoid running the instrument in
areas with ambient temperatures
above 25°C. Do not use the
instrument until you call (888) MJ
CYCLE (in the US or Canada), or
your local distributor (outside the
US or Canada).
15
PSSF
An irregularity was detected in
a power supply temperature
sensor. The sensor will be
checked and you will be
notified if any problem is found.
No action required at this time.
12
SBC-CAF
The block took slightly longer
than expected to achieve the
programmed temperature. This
condition may be caused by
inadequate airflow, or a
problem with the Alpha unit
itself.
Please observe the following
guidelines: 1) place instruments
at least 10cm apart, 2) avoid
placing instruments in any
location where hot air might
enter the intake vents, 3) place
instrument on a hard surface with
no debris or paper underneath,
4) clean all air vents and Alpha
unit fins of dust and debris, 5)
avoid running the instrument in
areas with ambient temperatures
above 25°C. Continue to use the
instrument. If problem reoccurs
call (888) MJ CYCLE (in the US
or Canada), or your local
distributor (outside the US or
Canada).
13
SLC
The heated lid took slightly
longer than expected to reach
its target temperature. There
may be a problem with the lid
sensor or lid heater.
Call (888) MJ CYCLE (in the US
or Canada), or your local
distributor (outside the US or
Canada).
Summary of Contents for DNA Engine Dyad PTC-220
Page 8: ...1 4 DNA Engine Dyad Operations Manual ...
Page 14: ...2 6 DNA Engine Dyad Operations Manual ...
Page 38: ...4 18 DNA Engine Dyad Operations Manual ...
Page 78: ...DNA Engine Dyad Operations Manual 5 40 ...
Page 109: ...DNA Engine Dyad Operations Manual 8 10 ...
Page 113: ...9 4 DNA Engine Dyad Operations Manual ...
Page 127: ...DNA Engine Dyad Operations Manual 10 14 ...
Page 136: ...11 9 Alpha Units and the Remote Alpha Dock System Figure 11 5 Flush mounting template ...
Page 137: ...11 10 DNA Engine Dyad Operations Manual ...
Page 142: ......
Page 156: ...DNA Engine Dyad Operations Manual In 12 ...
Page 157: ...Declaration of Conformity DoC 1 ...