Probes and Biopsy 13-13
Low-level disinfectant
EPA
Registration No.
Name
9480-6
SANI-CLOTH PLUS GERMICIDAL DISPOSABLE CLOTH
70627-56
OXIVIR TB
9480-9
PDI Sani-Cloth AF3 Germicidal Disposable Wipe
46781-8
CAVIWIPES
46781-6
CAVICIDE
13.1.6 Environment
Working condition
Use the probes in the following ambient conditions:
Probe
Ambient
temperature
Relative humidity
Atmospheric
pressure
L12-3E
0
℃
~40
℃
30%
~
85%RH (no
condensation)
700hPa~1060hPa
C11-3E
0
℃
~40
℃
30%
~
85%RH (no
condensation)
700hPa~1060hPa
SP5-1E
0
℃
~40
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
DE11-3E
18
℃
~30
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
SC6-1E
0
℃
~40
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
V11-3HE
0
℃
~40
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
L9-3E
0
℃
~40
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
SD8-1E
18
℃
~30
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
L14-5WE
0
℃
~40
℃
20%~85%RH (no
condensation)
700 hPa ~ 1060 hPa
P7-3E
10
℃
~40
℃
30%
~
85%RH (no
condensation)
700hPa~1060hPa
L20-5E
0
℃
~35
℃
15%~80%RH (no
condensation)
700 hPa ~ 1060 hPa
SC5-1NE
0
℃
~40
℃
20%
~
85%RH (no
condensation)
700hPa~1060hPa
Summary of Contents for DC-80A
Page 2: ......
Page 24: ......
Page 44: ......
Page 59: ...System Preparation 3 15...
Page 67: ...System Preparation 3 23...
Page 68: ......
Page 80: ......
Page 299: ...Probes and Biopsy 13 19...
Page 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Page 324: ......
Page 334: ......
Page 340: ......
Page 348: ......
Page 352: ......
Page 363: ...Barcode Reader B 11...
Page 368: ......
Page 382: ......
Page 391: ...P N 046 014137 00 3 0...