5-52 Image Optimization
3. Press
to enter the Smart 3D imaging preparation status, and define the ROI as well as the
curved VOI.
4. Select a render mode.
5. Select an acquisition method, and set the corresponding parameters on the touch screen.
Rocked mode: set [Angle] parameter.
Linear mode: set [Distance] parameter.
6. Press <Update> on the control panel to start the 3D image acquisition.
The system enters the 3D image viewing status when the acquisition is complete. Or, end the
acquisition by pressing <Freeze> or <Update> on the control panel.
In image viewing status, operations such as VOI setting and image editing can be performed. For details,
see “5.10.3.3 Static 3D Image Viewing.”
7. Exit Smart 3D.
Press <Update> or <Freeze> to return to Smart 3D image acquisition preparation. Or, press <B> or
to enter B mode.
5.10.5.2 Smart 3D Acquisition Preparation
In Smart 3D mode, you need to set the scan method in addition to the other items in Static 3D mode.
The other parameters are the same as those for Static 3D. For details, see “5.10.3.2 Static 3D
Acquisition Preparation.”
Method
Capture images using Linear scan or Rocked scan.
Linear scanning
Move the probe across the surface. See the following figure.
Rocked scanning
Rotate the probe once from the left to the right side (or from the right to the left) to include the
entire desired region. See the figure.
Summary of Contents for DC-80A
Page 2: ......
Page 24: ......
Page 44: ......
Page 59: ...System Preparation 3 15...
Page 67: ...System Preparation 3 23...
Page 68: ......
Page 80: ......
Page 299: ...Probes and Biopsy 13 19...
Page 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Page 324: ......
Page 334: ......
Page 340: ......
Page 348: ......
Page 352: ......
Page 363: ...Barcode Reader B 11...
Page 368: ......
Page 382: ......
Page 391: ...P N 046 014137 00 3 0...