System Preparation 3-5
Check the system after it is powered on
Check after the system is powered on:
No.
Check Item
1
There are no unusual sounds or smells indicating possible overheating.
2
There are no persistently displayed system error messages.
3
There is no evident excessive noise, or discontinuous, absent or black items in the B mode
image.
4
Check whether there is abnormal heat on the surface of the probe during an ultrasound
procedure. If you use a probe which is giving off excessive heat, it may burn the patient.
5
The control panel keys and knobs are fully functional.
6
The touch screen and the main monitor screens display normally depending on the system
modes and image status.
7
The exam date and time are the same as the system date and time, and are displayed
correctly.
WARNING:
If you use a probe giving off excessive heat, it may burn the patient.
If you find anything not functioning properly, this may indicate that
the system is defective. In this case, shut down the system
immediately and contact Mindray Customer Service Department or
sales representative.
3.2.5
Power the System OFF
You must follow the correct procedures to power the system off. Also, after you upgrade the
software or when the system is down, you need to power off and restart it.
To power the system off normally
Press the power button on left of the control panel to see the option:
Shut down: to power the system off normally.
Standby: to enter standby status.
Cancel: to cancel the operation.
NOTE:
1. DO NOT rush direct shutdown of the system. It may damage the data.
2. After the system is upgraded, use “Shut Down” to power the system off to make the
upgraded data effective.
Summary of Contents for DC-80A
Page 2: ......
Page 24: ......
Page 44: ......
Page 59: ...System Preparation 3 15...
Page 67: ...System Preparation 3 23...
Page 68: ......
Page 80: ......
Page 299: ...Probes and Biopsy 13 19...
Page 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Page 324: ......
Page 334: ......
Page 340: ......
Page 348: ......
Page 352: ......
Page 363: ...Barcode Reader B 11...
Page 368: ......
Page 382: ......
Page 391: ...P N 046 014137 00 3 0...