G-2 Electrical Safety Inspection
G.2
Device Enclosure and Accessories
G.2.1 Visual Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals, etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
G.2.2 Contextual Inspection
Test Item
Acceptance Criteria
The enclosure and accessories
No unusual noises (e.g., rattles inside the case).
No unusual smells (e.g., burning or smoky smells,
particularly from ventilation holes).
No taped notes that may suggest device
deficiencies or operator concerns.
G.3
Device Labeling
Check that the labels provided by the manufacturer or the healthcare facility are present and legible.
Main unit label
Integrated warning labels
G.4
Protective Earth Resistance
a. Plug the analyzer probes into the device’s protective earth terminal and the protective earth
terminal of the AC power cord.
b. Test the earth resistance with a current of 25 A.
c. Verify the resistance is less than the limits.
LIMITS
ALL COUNTRIES R = 0.2 Ω Maximum
Summary of Contents for DC-80A
Page 2: ......
Page 24: ......
Page 44: ......
Page 59: ...System Preparation 3 15...
Page 67: ...System Preparation 3 23...
Page 68: ......
Page 80: ......
Page 299: ...Probes and Biopsy 13 19...
Page 304: ...13 24 Probes and Biopsy NGB 035 NGB 039...
Page 324: ......
Page 334: ......
Page 340: ......
Page 348: ......
Page 352: ......
Page 363: ...Barcode Reader B 11...
Page 368: ......
Page 382: ......
Page 391: ...P N 046 014137 00 3 0...