– LOGOS ONE – Operator Manual
91
AFTER THE 3 MINUTES WITHOUT POWER, WITH UPS
The system automatically puts the specimens rack in “Safe mode”, according to the following schedule:
PHASE
CONDITION
POWER FAILURE
PROCEDURE
PHASE1
POWER FAILURE
PROCEDURE
PHASE2
POWER FAILURE
PROCEDURE
PHASE3
Fixation
Flushing
Rinsing1/2
Dehydration
Clearing
Wax
Cleaning
MAIN CAVITY
LOADING
FINISH LOADING
N.A.
N.A.
Fixation
Flushing
Rinsing1/2
Dehydration
Clearing
Wax
Cleaning WAX
PHASE
RUNNING
STOP THE PHASE
N.A.
N.A.
Cleaning
MAIN CAVITY
PHASE
RUNNING
STOP THE PHASE
TIMED
UNLOADING
N.A.
Fixation
WAX
UNLOADING
THE SOLUTION
FINISH
UNLOADING
LOADING
SAME
REAGENT
N.A.
Flushing
Rinsing1/2
Dehydration
Clearing
UNLOADING
THE SOLUTION
IF NEXT PHASE
EXISTS AND IS
NOT WAX
IF NEXT PHASE
EXISTS AND IS
WAX AVAILABLE
IF NEXT PHASE
EXISTS AND IS
WAX
NOT
AVAILABLE
FINISH
UNLOADING
FINISH
UNLOADING
FINISH
UNLOADING
LOADING
NEXT
REAGENT
LOADING WAX IN
MAIN CAVITY
RELOADING
SAME REAGENT
Cleaning
MAIN CAVITY
UNLOADING
THE SOLUTION
FINISH
UNLOADING
N.A.
N.A.
Since WAX cannot be melted in “Power failure procedure”, operator is required to wait till
the completion of WAX remaining heating time, before removing the specimens.
In case of power failure during the WAX impregnation in MAIN CAVITY or in WAX CAVITY,
the rack cannot be unloaded until the WAX heating time is completed (by default 100
minutes melting time). The specimens are safe anyway in the solid WAX, so there is no
risk of damage if they are not removed.
In case of power failure during the transfer of WAX between the two cavities, it is
impossible to start a new process until the WAX heating time is completed (by default 100
minutes melting time).