2.6.3
Main power supply connection
ONLY FOR 115V VERSION, add plug to the cable:
Connect the two black wires to power pins of
the plug and the yellow/green to ground’s pin.
Diameter approx. 10mm, size of wires AWG12.
ONLY FOR 230V VERSION, connect the cable to connector 1 (see the chapter 2.6.1).
2.6.4
Power supply connection (NO UPS mode)
There are two ways to connect the LOGOS ONE to power supply line.
Characteristic
: This is the easier way to supply the LOGOS ONE, but it does NOT protect the unit from
power supply failure.
In fact, in case of power failure, user cannot remove samples from LOGOS ONE nor do any other
operations.
For this reason, Milestone s.r.l. does NOT suggest this type of connection.
Make sure to inform the customer that in this way a power supply failure inhibits any operation with
LOGOS ONE
NEVER DIRECTLY CONNECT LOGOS ONE TO THE SWITCHBOARD.
IT IS REQUIRED TO USE A SEPARABLE PLUG, WITHOUT A LOCKING DEVICE to mate
with the socket-outlet in the laboratory
.
As separable power cord, it is strictly required to use power cord with wire size at least of
1.5mm² (equal to minimum AWG12). A size less of 1.5mm² (AWG12) cannot guarantee a
proper functionality of the unit and overheating of the power cord.