1.2 Designated uses
The Milestone LOGOS ONE and its accessories have been engineered for laboratory use and can therefore
withstand the harshest laboratory conditions for continuous use in Histological sample preparation
procedures.
The instrument is intended for laboratory use (indoor use) only.
The Milestone LOGOS ONE is specifically designed to automatically (when an automatic run is selected) or
semi-automatically (when a dual run is selected) execute these processes for histopathological applications:
Fixation, Post-fixation
Dehydration and Clearing
Drying
Paraffin wax infiltration
The Milestone LOGOS ONE must be operated exclusively with the reagents listed in section Compatible
reagents 1.7.
1.3 Technical specifications
CODE
61500/R:
Power supply input:
230V~ 50Hz 2000W,
CODE
61501/R:
Power supply input:
230V~ 60Hz 2000W.
Working Temperature: 15°C
– 30°C (59°F – 86°F)
Humidity (relative):
up to 80%
Maximum altitude:
2000m
Fan air flow range:
54
–58m
3
/h (cube meter per hour).
1.3.1
Touch control terminal
The following picture shows the Touch Control terminal of the LOGOS ONE unit. It is possible to operate the
unit by simply touching the icons on the screen.
Two USB PORTS are located on the right side of the Touch Control Terminal.
Please refer to chapter 9.4.1 for purity grade of reagents.
Any other use of the instrument is considered improper and may void the warranty