Wait until the new solid wax you have loaded has melted (100 minutes). If molten wax is loaded, the Wax
status icon becomes yellow. Check the wax status (fully molten) and press “Set as ready” button.
2.13.1 Wax heating setting
From the “Settings” screen select the “Heating” button (left figure below) or from the “Main menu” select the
WAX icon on the status bar (right figure below).
The “heating settings” screen appears displaying information about the Heating Status and the Cavity Level.
(Main processing chamber on the left and WAX chamber on the right in the same screen).
After the replacement, the software resets wax loading date automatically to the current
date and turns WAX HEATING ON.
By default WAX reuse life is checked by number of days (maximum 14 days). If you want
to change it, follow the instructions reported in the chapter 4.1.1.
MAX
MIN