Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
67
CASE 2
:
-
HistosMATE available to clean the rack
only 2 cleaning tanks required AND
-
Specimens to be processed include only little samples
In this case it is suggested to choose:
GROUP WITH 2 RINSINGS
TANK NR.
TANK REAGENT
REAGENT
1
FORMALIN
FORMALIN
2
FLUSHING MIX
ETHANOL from 60% to 70%
3
RINSING 1
ETHANOL 100%
4
RINSING 2
ETHANOL 100%
5
ETHANOL
ETHANOL 100%
6
ISOPROPANOL
ISOPROPANOL 100%
7
ISOPROPANOL (spare)
ISOPROPANOL 100%
8
CLEANING ISOPARAFFIN or XYLENE ISOPARAFFIN or XYLENE
9
CLEANING ETHANOL
ETHANOL from 96% to 100%
CASE 3
:
-
HistosMATE NOT available to clean the rack
3 cleaning tanks required AND
-
Specimens to be processed include big routine samples
2 isopropanols are required in overnight run
In this case it is suggested to choose:
GROUP WITH 1 RINSING
TANK NR.
TANK REAGENT
REAGENT
1
FORMALIN
FORMALIN
2
FLUSHING MIX
ETHANOL from 60% to 70%
3
RINSING 1
ETHANOL 100%
4
ETHANOL
ETHANOL 100%
5
ISOPROPANOL
ISOPROPANOL 100%
6
ISOPROPANOL 2
ISOPROPANOL 100%
7
CLEANING ISOPARAFFIN or XYLENE ISOPARAFFIN or XYLENE
8
CLEANING ETHANOL
ETHANOL from 96% to 100%
9
CLEANING WATER
WATER