3.2.5
Run a DELAYED PROGRAM
The window below appears, if either a Favorite program (see chapter 3.2.1) or a Standard program (see
chapter 3.2.2) is selected. Enter the exact number of cassettes in the box, choose if fixation phase must be
excluded or not, select th
e Run Mode and press “
Go delay
”.
The following window appears: set the completion time. In fact, “Delay Start” function allows setting when the
program has to finish, without modification of the existing protocol. The system keeps samples immersed in:
•
Pre-soaking reagent (formalin at room temperature)
OR
•
Holding reagent (flushing mix at room temperature)
By default Pre-soaking reagent is selected, if you want to keep the specimens immersed in Holding solution,
select its icon.
If you leave selected “Pre-soaking” (see above that is thicked by the arrow
) as reagent to
keep the specimens: the “Holding time” is zero.
If you press the grey arrow (figure above, yellow circle), the fixation protocol details are displayed:
•
“
Pre-soaking time
”: it depends on the starting time to complete the processing as set by user.
•
“
Fixation time at temperature
”: the duration of the heated fixation phase of the protocol.
•
“
Total fixation time
”: the sum of the two times above.
If you select
“Holding” (see below that is thicked by the arrow
) as reagent to keep the
specimens:
See the chapter 4.1.1 to set Pre-soaking and Holding reagent.