Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
31
Only when the WAX is ready for use (fully molten), the graph shows the lower level sensor as active
(green/covered) and the maximum level sensor as inactive (red/uncovered), as shown in the figure above.
OTHER FUNCTIONS:
The Molten Time is a pre-set value. Milestone s.r.l. strongly recommends NOT to modify it.
The maintaining temperature is a default value set according to Milestone s.r.l. standard
programs. The modification of this value is only allowed under direct control of Milestone
applicative support.
LOGOS ONE software will indicate with an alarm if the WAX level is incorrect. See
paragraphs 2.13 and 3.5 for how to manage the WAX, its reuse life and when proceeding
with the WAX cleaning protocol.
ADMINISTRATOR/SERVICE can modify all parameters of these fields; incorrect use of
these settings may lead to system malfunction. Use only under direct control of
Milestone s.r.l. technical support.
The
WAX heating status icon
indicates through a change in color
the:
WAX levels.
WAX “Temperature to maintain”.
The
reagent status icon
indicates status of WAX (and all the other reagents) and shows through a change
in color
the reagent status based on n° of cassettes processed and/or cycles
performed and/or expiration date. Please refer to chapter 4.
USER
can only “Set as ready” the WAX (if allowed by the ADMINISTRATOR in the USER
Settings) when the WAX heating status icon becomes YELLOW and the “Set as ready”
button becomes active (see the figures below).
It is only possible to “Set as ready” the WAX when its temperature is 4°C lower than the
preset temperature value, so a process can be immediately started.
In any case, it is necessary that operator checks actual status of the WAX inside the
CAVITY before setting “wax as ready”.
WAX is kept at this temperature when ON
Time to completely melt the WAX
WAX heating time remaining before use
WAX CAVITY stirrer speed
Restore WAX default settings