Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
34
Pipe number 1, 2, 4, 5, 7, 8 are provided with a RAPID CONNECTOR to simplify insertion/extraction. Just
press the rapid connector to easily insert/extract any pipes from each tank, as shown below.
Pipe number 3, 6 and 9 are not provided with rapid connectors, they are the longest pipes with the possibility
to simplify insertion/extraction without complete detaching of the pipe.
Press the rapid connector corresponding to the tube number 1, as shown below.
Choose the plastic ring suitable for the reagent you want to use in each position (in case of position
Nr. 1, Formalin is suggested)
as follows: press “Reagents list” from the Reagents configuration
screen, look to the color and the short name of the reagent (Formalin), then choose the
corresponding plastic ring.
The PLASTIC RING that corresponds to Formalin reagent (position 1) is white with the short name “FOR”.
Now insert it in the metallic pipe.
There are numbered stickers 1-9 on each pipe and their position.
It’s suggested to insert the Formalin tank in position number 1 and the Flushing tank in
position number 2.
TANK POSITION IS strongly suggested only for formalin and flushing.
Other tanks can be positioned according to user’s preference request.
PRESS
HERE