Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
45
Position the CASSETTE FIXING DISK on top of the last layer in use and load the rack into the LOGOS ONE.
3.1.1
Rack loading-unloading tool
Recommended to unload the rack at the end of the processing cycle, to avoid dripping of melted paraffin.
Recommended also to load rack with cassettes when it is wet, in order to avoid dripping of formalin or other
solutions on the floor or on the LOGOS ONE unit.
The first rack layer must be blocked in that position:
Do not insert the layer from the column: this will avoid stirrer rotation.
The fixing disk must always be used, with either 1, 2 or full layer rack in order to avoid
possible spilling out of cassettes from the rack during the processing.
If you position paper between the rack and the tool to keep it clean, remember to don’t
leave the rack dirty by wax up to the paper for many minutes because in that case the
paper will be attached to the rack.