3.3 Run a cleaning cycle of the main cavity
At the end of an automatic process (see the chapter 3.2.1.1), it is required to carry out a MAIN CAVITY
CLEANING, as this has been filled with WAX.
Unload the reagent and extract the rack by pressing the corresponding icons in the figure below (yellow
circles); download the cassettes and start the cleaning protocol in the processing screen, press START
CLEANING and run directly a cleaning program.
The CLEANING REAGENTS set in the Group of Programs page (SERVICE ONLY) are the only one that are
visualized in the cleaning page.
Choose the default program according to the cleaning reagents loaded in the REAGENT CONFIGURATION:
Only one program can be chosen by default and will be the one that starts automatically after an automatic
run that moves the paraffin in the main cavity. All the calculation for cavity availability are done on that cycle
duration.
The CLEANING REAGENTS set in the Group of Program page are the only one that are visualized in the
cleaning page.
Only one program can be chosen by default and will be the one that starts automatically after an automatic
run that moves the paraffin in the main cavity.
The same for the Wax Cleaning cycle, only one is set by default and all the calculation for wax and
chambers availability are done on that cycle.