Milestone s.r.l.
MM098-002
– LOGOS ONE – Operator Manual
12
2 INSTALLATION
Milestone LOGOS ONE is a precision instrument that requires the utmost care when it is unpacked and
installed. Move LOGOS ONE to its final location. Make sure that the floor is level and is made of non-
flammable material. The following pictures show the dimensions of LOGOS ONE and the space required:
2.1 Space requirements
LOGOS ONE is delivered in a wood box with these dimensions:
Width: 920mm (36.2in)
Height: 1320mm (52in)
Depth: 1500mm (59in)
The space dimensions required to locate the unit are the following (drawer and covers opened):
Width: 1100mm (43in)
Height: 1600mm (63in)
Depth: 1200mm (47in)
LOGOS ONE is a heavy instrument. Contact a structural engineer before placing the unit (analyze the
building structure).
LOGOS ONE Weight (read carefully):
LOGOS ONE FULL (all tanks): 320kg (706lb) (wax and all reagents loaded in proper tanks).
LOGOS ONE EMPTY: 250kg (552lb) (no reagents; no Paraffin Wax histological grade).
Unit density 640kg/m
2
(0.9lb/in
2
).
1500mm
1090mm
695mm
760mm