D39303 | Date 09/2018
7
For your safety
1.2
Product safety and dangers
The appliances described in this manual are technically sophisticated, manufactured using
high-quality materials and subject to many hours of testing in the factory. They reflect the
state of the art and comply with recognised technical safety regulations. However, there are
still risks involved, even when the appliances are used as intended. These are described below.
WARNING
After removing covers, live parts may be exposed. Touching these
can lead to an electrical shock. Disconnect the mains plug before
removing any covers. Work on the electrical system must only be
performed by quali
fi
ed electricians.
WARNING
When loading the appliance with an unsuitable chamber load, poi-
sonous or explosive vapours or gases may be produced. This could
cause the appliance to explode, and people could be severely injured
or poisoned. The appliance may only be loaded with materials/test
objects which do not release any toxic or explosive vapours when
heated (see also chapter Intended use and improper use on page
8).
WARNING
Hot steam can build up inside the appliance. You could be scalded
on opening the door. Allow the appliance to cool before opening the
door.
WARNING
Depending on operation, the surfaces in the interior of the appliance
and the chamber load may still be very hot after the appliance is
switched off. Touching these surfaces can cause burns. Wear heat-
resistant protective gloves or wait until the appliance cools down
before touching.
WARNING
Leaving the door open during operation can cause the appliance to
overheat or pose a
fi
re hazard. Do not leave the door open during
operation.
WARNING
Condensation in the electrical components may cause short circuits.
After transporting or storing the device under humid conditions, re-
move it from its packaging and let it ventilate for at least 24 hours in
normal environmental conditions. Do not connect the device to the
mains power during this time.