D39303 | Date 09/2018
39
Malfunctions, warning and error messages
Description
Cause
Action
See
Error message
T:E-3
in
the temperature display
TEMP
T:E-3
Set 37.0
°C
37
.4
°C
Temperature operating sen-
sor is defective. The moni-
toring sensor takes over the
measurement function.
►
The appliance can
temporarily be
kept in service
►
Contact customer
service as soon as
possible
Page 2
Error message
AI E-3
in
the temperature display
TEMP
AI E-3
Set 37.0
°C
37
.4
°C
Temperature monitoring
sensor is defective. The
operating sensor takes over
the measurement function.
►
The appliance can
temporarily be
kept in service
►
Contact customer
service as soon as
possible
Page 2
Error message
E-3
in
the temperature display
TEMP
Set 45.0
°C
E-3
°C
Operating and monitoring
sensor defective
►
Switch off appli-
ance
►
Remove the cham-
ber load
►
Contact customer
service
Page 2
Error message
E-6
in
the humidity display
HUMIDITY
Set 50
.0%rh
%rh
E-6
Humidity sensor defective
►
No humidity con-
trol possible
►
Contact customer
service
Page 2
When switching on
the appliance, the start
animation is displayed
in another colour than
white
►
Cyan
:
Not enough storage
space on the SD card
►
Red
:
The system files could
not be loaded
►
Orange
: The
fonts and images could
not be loaded
Contact customer
service
Contact customer
service
Download the firm-
ware update from
memmert.com and
install it
Page 2
Page 2