48
D39303 | Date 09/2018
Menu mode
1. Press the activation key to the right of the
CALIB
display. The display is enlarged
and the temperature adjustment option
is automatically selected.
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
40.0
60.0
30.0
Temperature
Humidity
2. Press the confirmation key repeatedly,
until the compensation temperature Cal2
is selected.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
re
Cal1
Cal2
Cal3
60.0
30.0
40.0
3. With the turn control, set the compensa-
tion temperature Cal2 to 42 °C.
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
r
Cal1
Cal2
Cal3
60.0
30.0
42.0
4. Save the setting by pressing the confir-
mation key. The corresponding calibra-
tion value is automatically highlighted.
C
K
0,2
-
C
K
C
K
0,2
-
re
Cal1
Cal2
Cal3
60.0
30.0
42.0
0,1
+
5. Set the calibration value to 0.0 K and ac-
cept the setting by pressing the confirma-
tion key.
C
K
0,2
-
C
K
C
K
0,2
-
re
Cal1
Cal2
Cal3
60.0
30.0
42.0
0,0
6. Position the sensor of a calibrated refer-
ence instrument centrally in the appli-
ance's working chamber.
7. Close the door and, in manual mode,
adjust the set temperature to 42 °C.
21
.4°C
TEMP
42
.0
°C
Set
8. Wait until the appliance reaches the set
temperature and displays 42 °C. The ref-
erence instrument should display 43.6 °C.
42
.0°C
TEMP
Set 42
.0°C
43.6
°C