Malfunctions, warning and error messages
6.1.2 Humidity monitoring (only for appliances in the corresponding
con
fi
guration)
Description
Cause
Action
See
Error display symbol
HUMIDITY
Set 55
.0%rh
%rh
55
.4
Water tank
empty
Fill the water tank with deminer-
alised/distilled water and press the
confirmation key
Page
22
Alarm display (
MaxAl
)
HUMIDITY
MaxAl
Set 70
.0%rh
%rh
75
.4
Upper humidity
limit exceeded
Open the door for 30 sec. and
wait to see if the appliance reli-
ably adjusts to the setpoint.
If the error occurs again, contact
customer service.
Page 2
Alarm display (
MinAl
)
HUMIDITY
MinAl
Set 60
.0%rh
%rh
55
.4
Humidity below
lower limit
Check if the door is closed.
Check the water supply and the
filling level of the water tank. If
required, refill the water tank/
water trays.
If the error occurs again, contact
customer service.
Page
22
Page 2
6.2
Malfunctions, operating problems and appliance errors
Description
Cause
Action
See
Displays are dark
External power supply
was interrupted
Check the
power supply
Page 21
Miniature fuse, appliance
fuse or power module
faulty
Contact customer
service
Page 60
Displays cannot be
activated
Appliance locked by USER
ID
Unlock with USER ID
Page 54
The appliance is in pro-
gram, timer or remote
control mode (mode "Write"
or "Write + Alarm")
Wait until the end of
the program or timer
mode or switch off the
remote control
Displays suddenly look
different
Appliance is in "wrong"
mode
Change to operating
or menu mode by
pressing the
MENU
key