Operation and control
5. With the turn control, select ON (
) or
OFF (
).
ALARM
max
38
.5°C
0.0
K
min
35
.5°C
auto
+-
6. Press the confirmation key to confirm.
The ASF tolerance band setting is acti-
vated.
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
0.0
+-
7.
With the turn control, adjust the desired
tolerance band. We recommend 1 to 3 K.
ALARM
K
min
°C
auto
max
38
.
5°C
35
5
.
1.0
+-
8. Press the confirmation key to confirm.
Temperature monitoring is now active.
ALARM
max
38
.5°C
1.0
K
min
35
.5°C
auto
+-
5.6.2
Humidity monitoring
If humidity monitoring was triggered, this
is indicated by the humidity display: the
actual humidity is highlighted in red and a
warning symbol
is shown ( Fig. 19 ). If the
acoustic alarm has been activated in menu
mode (
Sound
, see page 52, as indicated
by the speaker symbol
), the alarm is
additionally signalled by an intermittent
acoustic signal. Information on what to
do if this happens can be found in chapter
Malfunctions, warning and error messages
from page 41.
Adjusting humidity monitoring
1. Press the activation key to the left of the
ALARM
display. The temperature moni-
toring setting is automatically activated.
min
000°C
ALARM
max
000
°C
auto off
99K
-
+
ALARM
max
40
.0°C
auto
0.0
K
min
15
.0°C
+-
4
m
min
000°
O
LIG
O
O
O
O
LIG
IG
IG
O
LIG
FEUCHTE
30
.0%rh
Set 30
.0%rh
0
°C
°C
°C
22
.4
EMP
°C
Set 37.0
a
m
0
m
3:30
HUMIDITY
Set 70
.0%rh
%rh
75
.4
m
FEUCHTE
30
.0%rh
Set 30
.0%rh
C
°C
Set 37.0
m
0
HUMIDITY
Set 70
.0%rh
%rh
75
.4
Fig. 19
Humidity monitoring triggered
Summary of Contents for HCP105
Page 1: ...HCP OPERATING MANUAL HUMIDITY CHAMBER HCP...
Page 59: ......