40
D39303 | Date 09/2018
Malfunctions, warning and error messages
6.3
Power failure
In case of a power failure, the appliance operates as follows:
In manual mode
After power supply has been restored, operation is continued with the parameters set. The
time and duration of the power failure are documented in the log memory.
In timer or program mode
In case of an interruption of the power supply of less than 60 minutes, the current program is
continued from the point at which it was interrupted. For longer interruptions of the power
supply, all appliance functions (heating, fan etc.) are switched off.
In remote control mode
The previous values are restored. If a program has been initiated via remote control, it is con-
tinued.