Menu mode
t
Timer
t
Timer
A
B
°C/°F
t
6 K
Fig. 23
Timer mode
A
Timer independent of setpoint: Timer starts immediately after activation
B
Timer setpoint-dependent: Timer does not start until tolerance band is reached
7.3.5
Remote control (AtmoREMOTE)
In the setup entry remote control, you can set whether the
appliance should be controlled via remote control and if so,
in which mode. These settings are available:
►
Off
►
Read only
►
Write
►
Write + Alarm
If the appliance is in remote control mode, the symbol
appears in the temperature display. In the settings
Write
and
Write + Alarm
, the appliance cannot be controlled at
the ControlCOCKPIT until the remote control has been
switched off (setting
Off
) or set to
Read only
.
In order to use the remote control function, program-
ming skills and special libraries are required.
7.3.6
Gateway
The setup entry gateway is used to connect two networks
with different protocols.
The gateway is set the same way as the IP address (see page
43).
7.4
Date and time
In the
TIME
display, you can set the date and time, time zone and daylight saving time.
NOTICE
Always set the time zone (and daylight saving time yes/no) before you set the date and
time. Avoid changing the set time after that since this can lead to gaps or overlapping when
recording measured values. If you still need to change the time, you should not run a program
immediately before or after doing so.
Remote Control
Off
Gateway 192.168.5 .1
23
.2°C
TEMP
Set 38
.0°C
Remote Control Off
Gateway
192.168.5 .1