44
D39303 | Date 09/2018
Menu mode
1. Activate the
SETUP
display. The entry
IP
address
is automatically highlighted.
SETUP
Balance
Einheit
IP Adresse
+30
%
192.168.100.100
Alarm Temp
°C
F
TWW
TWB
Timer Mode
SETUP
Unit
IP address
192.168.100.100
Subnet mask
255.255.0.0
Alarm Temp
°C
F
TWW
TWB
Timer Mode
2. Accept the selection by pressing the con-
firmation key. The first three digits of the
IP address are automatically selected.
SETUP
Einheit
IP-Adresse
192
.168.100.100
Subnetmask
255.255.0.0
Alarm Temp
°C
F
TWW
TWB
Timer Mode
3. With the turn control, set the new num-
ber, e.g. 255.
SETUP
Unit
IP address
255
.168.100.100
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
4. Accept the selection by pressing the
confirmation key. The next three digits of
the IP address are automatically selected.
Setting these is done according to the
description above.
SETUP
Unit
IP address
255.
168
.100.100
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
5. After setting the last three digits, accept
the new IP address by pressing the con-
firmation key. The selection returns to the
overview.
The subnet mask is set accordingly.
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode
7.3.3
Unit
Here you can choose whether the temperature is displayed in
°
C
or °F.
7.3.4
Timer Mode
Here, you can choose whether the digital backwards counter with
target time setting (timer, see page 27) should run setpoint-
dependent or not – this determines whether the timer should not
start until a tolerance band of ±3 K around the setpoint tempera-
ture is reached (Fig. 23, B) or whether it should start right after
activation (A).
SETUP
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
°F
TWW
TWB
Timer mode
Unit
IP address
255.145.136.225
Subnet mask
255.255.0.0
Alarm temp
°C
F
TWW
TWB
Timer mode