D39303 | Date 09/2018
55
Maintenance and Servicing
8.
Maintenance and Servicing
WARNING
Danger due to electric shock. Disconnect the mains plug before any
cleaning or maintenance work.
8.1
Regular maintenance
Annually:
►
Check the sterile filter in the control unit and the water pump filter in the rear panel and
replace them if they are dirty.
►
To guarantee perfect control, we recommend calibrating the appliance once a year (see
page 47).
Every two years:
►
Replace the sterile filter in the control unit and the water pump filter.
8.2
Cleaning
8.2.1 Interior and metal surfaces
Regular cleaning of the easy-to-clean interior prevents build up of material remains that could
impair the appearance and functionality of the stainless steel chamber over time.
The metal surfaces of the appliance can be cleaned with normal stainless steel cleaning
agents. Make sure that no rusty objects come into contact with the interior or with the stain-
less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots
should appear on the surface of the interior due to impurities, the affected area must be im-
mediately cleaned and polished.
8.2.2 Plastic parts
Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic or
solvent-based cleaning agents.
8.2.3 Glass surfaces
Glass surfaces can be cleaned with a commercially available glass cleaner.
8.3
Repairs and Service
Only authorised customer service points should carry out maintenance work.