Menu mode
10. In the SETUP, adjust the compensation
correction value Cal2 to -1.5% (actual
value measured minus setpoint humid-
ity) and save the setting by pressing the
confirmation key.
%rh
%rh
%
0,5
-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0
-1.5
%rh
%
1,0
+
Cal3
80.0
11. After the calibration procedure, the hu-
midity measured by the reference instru-
ment should now also be 60% rh.
HUMIDITY
60
.0%rh
Set 60
.0%rh
60.0
%rh
7.6
Program
In the
Program
display, programs created using the AtmoCONTROL software can be trans-
ferred to the appliance and saved on a USB storage medium. Here, you can also select the
program provided for use (see page 28 ) and delete programs.
To load a program from a USB stor-
age medium: Connect the USB storage
medium with the saved program(s) to the
interface on the right side of the Control-
COCKPIT.
1. Activate the program display. To do so,
press the activation key on the left side of
the
Prog
display. The display is enlarged
and the entry
Select
automatically
highlighted. The programs available for
activation are shown on the right. The
program currently available for use – in
this example
Test 012
– is highlighted in
orange.
Program
Delete
Test 012
Test 013
Test 014
Test 015
Test 016
Select
2. Access the
Select
function by pressing
the confirmation key. All programs avail-
able are displayed, including the ones
saved on the USB data storage medium
(identified by the
USB symbol). The
program currently available for use is
highlighted in orange.
Select
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 022
3. With the turn control, select the program
you want to make available for use.
Select
Delete
Test 012
Test 022
Test 013
Test 014
Test 015
Test 023
Summary of Contents for HCP105
Page 1: ...HCP OPERATING MANUAL HUMIDITY CHAMBER HCP...
Page 59: ......