D39303 | Date 09/2018
21
Putting into operation
4.
Putting into operation
NOTICE
When putting the appliance into operation for the first time, do not leave it unattended until
it has reached a steady state.
4.1
Connect the appliance to the power supply
WARNING
Condensation in the electrical components may cause short circuits.
After transporting or storing the device under humid conditions, re-
move it from its packaging and let it ventilate for at least 24 hours in
normal environmental conditions. Do not connect the device to the
mains power during this time.
Observe the country-specific regulations when making
connections (e.g. in Germany DIN VDE 0100 with earth
leakage circuit breaker). Observe the connection and
power ratings (see nameplate and "Technical Data" on
page 14). Make sure to establish a safe PE conductor
connection.
Plug the provided power cable into the rear of the appli-
ance and connect it to the power supply ( Fig. 10 ). Place
the power cable so that
►
it is easily accessible at all times and can be pulled off quickly, for example in case of inter-
ference or an emergency;
►
it does not represent a trip hazard;
►
it cannot come into contact with any hot parts.
4.2
Establish water supply
4.2.1
Water speci
fi
cations
For steam creation, use only: demineralised/desalinated water (Aqua Dem) with a conduc-
tivity
≤
10
μ
S/cm. Otherwise, possible calcification in the steam generators, steam piping
and steam generator pumps could affect the correct functioning of the appliance. Only use
chlorine-free water with a pH value of > 5 and < 7.
Fig. 10
Connect the power
cable to the rear of the appliance