50
D39303 | Date 09/2018
Menu mode
2. Turn the turn control until
Humidity
is
highlighted.
%rh
%rh
%
0,5
-
%
1,0
+
Temperature
Cal1
Cal2
40.0
50.0
Humidity
%rh
%
1,0
+
Cal3
80.0
3. Press the confirmation key repeatedly,
until the calibration point Cal2 is selected.
%rh
%rh
%
0,5
-
%
1,0
+
Temperature
Cal1
Cal2
40.0
Humidity
50.0
%rh
%
1,0
+
Cal3
80.0
4. With the turn control, set the calibration
point Cal2 to 60% rh.
%rh
%rh
%
0,5
-
%
1,0
+
Temperature
Cal1
Cal2
40.0
Humidity
60.0
%rh
%
1,0
+
Cal3
80.0
5. Save the setting by pressing the confir-
mation key. The corresponding calibra-
tion value is automatically highlighted.
%rh
%rh
%
0,5
-
%
JUSTIEREN
Temperature
Cal1
Cal2
40.0
Humidity
60.0
1,0
+
%rh
%
1,0
+
Cal3
80.0
6. Set the calibration value to 0.0% and ac-
cept the setting by pressing the confirma-
tion key.
%rh
%rh
%
0,5
-
%
Temperature
Cal1
Cal2
40.0
Humidity
60.0
0,0
%rh
%
1,0
+
Cal3
80.0
7. Position the sensor of the calibrated refer-
ence instrument centrally in the working
chamber of the appliance.
8. Close the door and, in manual mode,
adjust the set humidity to 60% rh.
HUMIDITY
29
.5%rh
Set
60
.0
%rh
9. Wait until the appliance reaches the set
humidity and displays 60% rh. The refer-
ence instrument should display 58.5% rh.
HUMIDITY
60
.0%rh
Set 60
.0%rh
58.5
%rh