Delivery, transport and setting up
3.6
Setting up
3.6.1 Preconditions
The installation site must be flat and horizontal and must be able to reliably bear the weight
of the appliance (see Technical data on page 14). Do not place the appliance on a flamma-
ble surface.
Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail-
able at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm. The clear-
ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby
appliances must not be less than 5 cm (Fig. 9). Sufficient air circulation in the vicinity of the
appliance must be guaranteed at all times.
FP
FP
FP
FP
Fig. 9
Minimum clearance from walls and ceiling