12
D39303 | Date 09/2018
Construction and description
2.4
Material
For the outer housing, MEMMERT uses stainless steel (Mat.No. 1.4016 – ASTM 430) and for
the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through
its high stability, optimal hygienic properties and corrosion resistance to many (but not all!)
chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with
the materials mentioned. A material resistance table can be requested from the manufacturer.
2.5
Electrical equipment
►
Operating voltage and current consumption: See nameplate
►
Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010
►
Protection type IP 20 acc. to EN 60 529
►
Interference suppression acc. to EN 55011 class B
►
Appliance fuse: Safety fuse 250 V/15 A, quick-blow
►
The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)
2.6
Connections and interfaces
2.6.1
Electrical connection
This appliance is intended for operation on an electrical power system with a system imped-
ance Z
max
of a maximum of 0.292 ohm at the point of transfer (service line). The operator
must ensure that the appliance is operated only on an electrical power system that meets
these requirements. If necessary, you can ask your local energy supply company what the
system impedance is.
Observe the country-specific regulations when making connections (e.g. in Germany DIN VDE
0100 with earth leakage circuit breaker).
2.6.2
Communication interfaces
The communication interfaces are intended for appliances which meet the requirements of
IEC 60950-1.
USB interface
The appliance is fitted by default with a USB
interface in accordance with the USB specifica-
tion. This way, you can
►
transfer software stored on a USB storage
medium to the appliance (see page 51).
►
export protocol logs from the appliance to
a USB storage medium (see page 53).
►
transfer user ID data stored on a USB stor-
age medium to the appliance (see page
54).
The USB port is located on the right side of the ControlCOCKPIT (Fig. 4).
Fig. 4
USB interface
Summary of Contents for HCP105