24
D39303 | Date 09/2018
Operation and control
5.4
Loading the appliance
WARNING
When loading the appliance with an unsuitable load, poisonous or
explosive vapours or gases may be produced. This could cause the
appliance to explode, and people could be severely injured or poi-
soned. The appliance may only be loaded with materials which do
not form any toxic or explosive vapours when heated up and can-
not ignite (see also Intended use and improper use on page 8). If
there is any doubt as to the composition of materials, they must not
be loaded into the appliance.
NOTICE
►
Check the chamber load for chemical compatibility with the materials of the appliance (see
page 12).
Insert the sliding steel grids or sliding shelves. The maximum number of grids/shelves and the
load capacity are specified in the technical data overview from page 14.
The appliance must not be
loaded too tightly, so that
proper air circulation in the
interior is guaranteed. Do not
place any of the chamber
load on the bottom, touching
the side walls or right below
the ceiling of the chamber
(Fig. 14, see also the "correct
loading" sticker on the
appliance).
In case of improper loading
(not enough space between
the items), the set tempera-
ture may be exceeded or it
may take longer until it is
reached.
5.5 Operating the appliance
5.5.1
ControlCOCKPIT
In manual mode, the desired parameters are entered in the ControlCOCKPIT on the front of
the appliance (Fig. 15 ). You can also make basic settings here (menu mode). Additionally,
warning messages are displayed, e.g. if the temperature is exceeded. In program mode, the
parameters defined, the program description, the program segment currently active and pro-
gram duration remaining are displayed (for a more detailed description, see page 28).
Fig. 14
Correct placement of the chamber load