Revision 1.7.2
Page 191
Figure 213
– Tank Levels Calibration Capacity Estimation
The next step in the calibration process is to fill the tank and indicate the volume of fluid in the tank at
various points while filling the tank. This process should be started with an empty tank. You only need
to indicate two calibration points for a rectangular tank, and can use up to 16 calibration points for a
tank with a more complex shape. Figure 214 shows a screenshot of the first calibration step. Enter
the first calibration point with an empty tank (0.0 Gallons/liters) by pressing the
Accept
key. The
“Current Point” highlight will advance to the next point.
Figure 214
– Empty Level Calibration
Each time you press
Accept
during this process and after the Current Point highlight advances, put
more fluid into the tank, and note the total volume of fluid pumped into the tank at this point. Use the
Increase
and
Decrease
keys to set the current tank level to the appropriate value, and then press the
Accept
key to store the current calibration point and advance to the next calibration point. Figure 215
shows a screenshot of these intermediate calibration points.