Part No. 20717 Rev-01
2014-07
3-18
Location of nebulizer distal from the inspiratory
gas sample tee
Disk Filter, 0.5 micron
Figure 3-9
Figure 3-10
Inspired Gas Sampling During Aerosol Delivery
• To avoid a “Sample Line/Filter Block” alarm
condition, place the medication nebulizer
downstream of the sample tee on the inspiratory
limb to avoid over saturation of the water
separator cartridge or contamination of the
sample system (see Figure 3-9).
• To minimize replacement of filters during aerosol
delivery, a small 0.5 micron hydrophobic disk filter
with luer connections must be placed between the
sample line and sample tee (see Figure 3-10).
Note:
The disk filter is to be used in
conjunction with the INOmax DS
IR
water
separator cartridge. The INOmax DS
IR
must never be operated without the
water separator cartridge.
• This disk filter has been validated for
this purpose.
• This disk filter must be replaced
following each treatment period.
• During continuous medication delivery,
frequent disk replacement may be
necessary due to the disk becoming
saturated with moisture/medication.
• Use of the filter will reduce the
frequency of water separator cartridge
replacement during medication
delivery.
Caution:
Pneumatic nebulizers will dilute the delivered INOMAX dose.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...