3. Connect the INOmax DS
IR
power cord to an
emergency-power-backed hospital-grade
outlet. The power cord must always be
connected to an electrical outlet to maintain
a full battery charge.
Verify the main power supply indicator light
is illuminated (green). This indicates that the
power cord is plugged into an electrical outlet.
3
Caution:
Keep the power cord off of the
ground and away from moving
parts.
4. Load two INOMAX drug cylinders onto
cart and check for correct product identity
labels, cylinder concentration (800 ppm) and
expiration date.
5. Connect an INOMAX regulator to one of the
INOMAX cylinders, and hand tighten the fitting
to the INOMAX cylinder. Connect the INOMAX
regulator hose to one of the INOMAX inlets
(see page 7-9 for additional instructions).
Caution:
If using the transport regulator/cap
assembly (PN 10022) see Figure
4-7, Section 4/ Transport.
Note: For the CGA-type INOMAX regulator
connector, ensure the white plastic tip
is not chipped or cracked. Remove and
replace as necessary. (see Replacing the
tip on the INOMAX regulator, Page 7-8).
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...