Part No. 20717 Rev-01
2014-07
1-26
The National Institute for Occupational Safety and
Health (NIOSH) have recommended exposure
limits as follows (Ref. 1).
NO
time-weighted (8 hours) average
concentration limit of 25 ppm
NO
2
ceiling limit of 1 ppm.
The environmental build up of NO in a well
ventilated ICU room can be evaluated using the
following calculation.
Room size
1000 ft
3
Room volume
28,300 L
Room ventilation
(6 complete exchanges/hour)
2,830 L/min
NO flow into the room
80 ppm at
14 L/min
Average NO room
concentration
(80 x 14) ÷ 2,830
(80 x 14) ÷ 2,830 = 0.396 ppm
(0.4 ppm)
0.4 ppm of NO
This theoretic calculation can be supplemented by
measurements as performed by Hess et al (Ref. 2).
The NO and NO
2
concentrations were measured
using a chemiluminescence analyzer when 100
ppm of NO at 8 L/min was delivered into a room
with no scavenging being used. The maximum NO
and NO
2
concentrations measured over a one hour
period were 0.12 ppm of NO and 0.03 ppm of NO
2
.
Environmental Effects
Both these methods show that the exposure levels
are significantly less than the levels recommended
by NIOSH.
If the location for using NO has uncertain
ventilation then the location should be evaluated for
NO and NO
2
build up prior to use.
References:
(Ref. 1) Centers for Disease Control, Atlanta, GA
30333 USA.
NIOSH Recommendations for Occupational
Safety and Health Standards 1988.
August 26, 1988 / vol. 37 / No. 9.
(Ref. 2) Hess et al, Use of Inhaled Nitric Oxide in
patients with Acute Respiratory Distress
Syndrome.
Respiratory Care, 1996, vol. 41, No. 5,
pg. 424-446.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...