Part No. 20717 Rev-01
2014-07
5-6
High Priority Alarms
Alarm
Possible Cause
Recommended Action
1. High NO alarm
All of these actions can be performed while
delivering INOMAX to the patient:
a. The High NO alarm level
may be inappropriately set.
Confirm high NO alarm limit is set appropriately.
b. Circuit setup incorrect.
Check circuit for correct setup.
c. The NO calibration may
have drifted.
1. Perform low calibration.
2. Perform high NO calibration.
d. Injector module may not be
functioning properly.
INOMAX delivery will be interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
1. Change injector module.
2. Change injector module cable.
“Injector Module Failure” alarm will occur
and NO delivery will be interrupted. This
will resolve when injector module cable is
reconnected.
Contact Technical Support.
Alarm Help
Continuous Audible Tone
Alarm
Possible Cause
Recommended Action
Continuous Audible Tone
A component within the
INOmax DS
IR
has failed.
1. If the INOblender is available, manually
ventilate the patient (see INOblender
Operation Manual).
2. Replace the delivery system and remove from
service.
3. Contact Technical Support.
WARNING:
• Abrupt discontinuation of INOMAX may lead to worsening oxygenation and
increasing pulmonary artery pressure, i.e., Rebound Pulmonary Hypertension
Syndrome. To avoid abrupt discontinuation, utilize the INOblender or integrated
pneumatic backup if necessary. If Rebound Pulmonary Hypertension occurs,
reinstate INOMAX therapy immediately. (See the INOMAX prescribing
Information for further details).
• If the high NO
2
alarm activates, the delivery system should be assessed for
proper setup while maintaining INOMAX delivery. Adjust INOMAX and/or FiO
2
as appropriate. (See INOMAX Prescribing Information for further details on
the effects of Nitrogen Dioxide, NO
2
). If unable to determine the cause of the
increased NO
2
levels, call technical support and do not discontinue therapy.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...