Part No. 20717 Rev-01
2014-07
9-6
Assemble connectors and tubing as shown in
Figure 9-1 to perform the following three procedures:
Figure 9-1
1. Verify the INOMAX cylinder valve is closed.
2. Set the O
2
flowmeter to 10 L/min (#1 in
Figure 9-1).
3. Purge INOmax DS
IR
.
• Set the INOMAX dose to 40 ppm.
• "Cylinder Valve Closed" alarm will occur.
• Cylinder gauge pressure should drop to
0 psig.
• Measured NO
2
will increase and then de-
crease as NO
2
is purged from the system.
• “Low Cylinder Pressure” alarm will occur.
4. Open the INOMAX cylinder valve.
5. Turn the INOMAX dose to zero. The “Set
Dose is Zero, Close Cylinder Valve” indicator
will appear. This indicator will display anytime
the set dose is returned to zero; however,
during this pre-use procedure, leave the
cylinder open and touch the screen to reset
the indicator.
Manual Purge and Alarm Verification
6
1
2
3
4
5
8
7
10
9
1. O
2
Flowmeter (Connected to wall/tank)
2. Injector Module Electrical Cable
3. NO/N
2
Injector Tube
4. Patient Gas Sample Line with Nafion
5. O
2
Tubing
6. 15M x 4.5 mm Adapter
7. 22M / 15F x 22M / 15F Adapter
8. Injector Module
9. 300 mm of 22 mm Hose
10. Gas Sample Tee
Note: Allow the auto low calibration to complete before
completing the manual purge and alarm verification.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...