Part No. 20717 Rev-01
2014-07
4-10
INOblender Stand-Alone Pre-use Checkout
High-Pressure Leak Test
1. Make sure NO dose setting dial is turned to
zero and flow meter is OFF.
2. Open and then close the INOMAX cylinder
valve.
3. Check the INOMAX drug cylinders for the
correct product identity, cylinder concentration,
and expiration date. Verify cylinder has at
least 500 psig (replace if 200 psig or less), and
tighten the fitting to the INOMAX cylinder.
4. Monitor pressure gauge for 30 seconds for
any signs of pressure decrease. If no pressure
decrease is observed, high-pressure leak test
successful, proceed to Delivery Confirmation
and Purge.
5. If observed pressure decrease continues, see
Section 7/ Maintenance; Cylinder Leak Check.
6. If leak cannot be traced, replace the
INOblender.
2
3
Caution:
To help ensure proper operation, complete the pre-use checkout prior to each use.
Delivery Confirmation and Purge
1. Set the INOblender to 40 ppm when using an
800 ppm cylinder.
2. Set the oxygen flow on the INOblender flow
meter to 10 L/min to begin purge.
3. Ensure the pressure gauge decreases
approximately
14 bar (200 psig) in 10 seconds
(+ two seconds).
4. Continue purging until pressure gauge
reads zero.
3
4
1
2
Note: If the pressure does not decrease, then
the INOblender is not delivering NO and
the INOblender should be replaced.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...