Part No. 20717 Rev-01
2014-07
8-6
Electrical
Input Voltage:
100-240 V AC @ 50 / 60 Hz
Input Power:
110 VA max
Input Fuse:
3 A
Classification:
Class I, Type B
Standards:
CSA certified to meet the following for medical electrical equipment:
• UL 60601-1: 2003 edition 2
• ANSI/AAMI 60601-1: 2005 edition 3
• IEC 60601-1: 2005 edition 3
Battery Backup: A sealed lithium ion rechargeable battery provides power backup
to operate the system for up to six hours when fully charged.
Connect the system to an electrical outlet for at least ten
hours to charge the battery.
When the low battery alarm occurs, there are 30 minutes
until battery depletion.
Dispose of used batteries according to local regulations.
USB Port:
Not used. Not for use when patient is connected.
Ethernet Port:
For service only. Not for use when patient is connected.
RS232:
Enables serial communications for use with hospital electronic health record
(EHR) system (see
specifications on page 8-7
).
Infrared Port:
Infrared communication with the INOMAX cylinder.
Important: Disconnect main power cord to isolate equipment from main power.
Environmental
Operating:
Transport/Storage:
Temperature:
5 to 40°C
-20 to + 60°C
Humidity:
15 to 95% RH
non-condensing
15 to 95% RH
non-condensing
Ambient
Pressure:
57 to 110 kPa
57 to 110 kPa
Water Ingress
Protection:
IPX1
INOMAX Regulator
Inlet Pressure:
14 to 155 Bar (200 to 2,248 psig)
Outlet Pressure:
1.7 to 2.4 Bar (25 to 35 psig)
Cylinder Valve Connector: CGA 626
Alarm Log
The alarm history is deleted when device is turned off. However, the service log, which is
accessible by service personnel is maintained (including alarm log) when power is cycled
and/ or when total power loss occurs.
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...