Part No. 20717 Rev-01
2014-07
3-52
Viasys Infant Flow SiPAP
• The INOmax DS
IR
adds NO/N
2
gas flow to the breathing circuit flow in proportion to the NO setting
(up to 10% at 80 ppm) and subtracts gas from the breathing circuit via gas sampling at a nominal
flow rate of 0.23 L/min.
• These effects change the flow going to the nasal adapter and can therefore impact the CPAP level
established by specific flow settings (See table below). The maximum flow error is approximately
11% at two L/min which is within the accuracy of the flow meter specification (+/-15%).
• It is recommended that after an NO setting change the user checks the CPAP level on the Infant
Flow SiPAP front panel display and adjusts as necessary.
Flow
(L/min)
Flow
(L/min)
Flow
(L/min)
Flow
(L/min)
Flow
(L/min)
SiPAP Flow Setting
2
4
6
8
10
After INOmax
DS
IR
Set @ 0 ppm
1.77
3.77
5.77
7.77
9.77
% error
-11.5%
-5.8%
-3.8%
-2.9%
-2.3%
After INOmax
DS
IR
Set @ 80 ppm 1.97
4.17
6.37
8.57
10.77
% error
-1.5%
4.3%
6.2%
7.1%
7.7%
Summary of Contents for INOMAX DSir Plus
Page 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Page 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Page 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Page 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Page 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Page 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Page 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Page 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Page 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Page 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Page 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Page 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Page 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Page 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Page 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Page 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Page 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Page 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...